Human Action Cycle for Set Phasers On Stun:
Goal: Give proper radiation treat to patient after
his surgery for removal of tumor
Intention: Use
Therac-25 cancer radiation therapy machine
Sequence of
Actions: Place patient underneath
machine
Go to
control room
Enter
commands into keyboard (e for electron beam)
Initiate
treatment by pressing b for turning beam on
Actions: Same as
above sequence, however she pressed “x” instead of “e” , pressed up, edits and
reenters “e” then follows to beam
ready, and presses b
World: Patient
hit by massive radiation, error appears on screen of display,
Perception: Mary
Beth sees the error on her screen but doesn’t see that patient is hit with
massive radiation
Interpretation:
Mary Beth thinks that machine hadn’t fired and tries to repeat the procedure
again. (does this 3 times), She thinks patient hasn’t received treatment yet
Evaluation:
proper radiation treatment has not been given to patient but Mary do not know
what has happened in the room. Her goal is not met
Marking notes:
Above is the best answer. Some may have different goals or intentions, as long
as the cycle is followed through you will get full mark. 1 mark for each step’s
label and description.